|
Calibration of a coaxial step attenuator at a setting of 30 dB (incremental loss)
This Example is taken from EA 4/02. See EA 4/02 Section S7 for more details.
The measurement involves the calibration of a coaxial step attenuator at 10 GHz using an attenuation measuring system containing a calibrated step attenuator which acts as the attenuation reference. The method of measurement involves the determination of the attenuation between a matched source and matched load. In this case the unknown attenuator can be switched between a setting of 0 dB and 30 dB and it is this change (called incremental loss) that is determined in the calibration process. The attenuation measuring system has a digital readout and an analogue null detector which is used to indicate the balance condition.
Model Equation:
| LX = LS + dLS + dLD + dLM + dLK + dLib - dLia + dL0b - dL0a
|
List of Quantities:
|
Quantity
|
Unit
|
Definition
|
|---|
| LX
|
dB
|
attenuation of the attenuator to becalibrated
|
| LS
|
dB
|
=Lib - Lia attenuation difference of reference attenuator at 30 dB and 0 dB
|
| dLS
|
dB
|
correction obtained from the calibration of the reference attentuator
|
| dLD
|
dB
|
change of the attenuation of the reference attenuator since its last calibration due to drift
|
| dLM
|
dB
|
correction due to mismatch loss
|
| dLK
|
dB
|
correction for leakage signals between input and output of the attenuator to be calibrated due to imperfect isolation
|
| dLib
|
dB
|
correction due to the limited resolution of the reference detector at 30 dB
|
| dLia
|
dB
|
correction due to the limited resolution of the reference detector at 0 dB
|
| dL0b
|
dB
|
correction due to the limited resolution of the null detector at 30 dB
|
| dL0a
|
dB
|
correction due to the limited resolution of the null detector at 0 dB
|
|
LS: |
Type A
Method of observation: Direct
Number of observation: 4
|
No.
|
Observation
|
|---|
| 1
|
30.033 dB
|
| 2
|
30.058 dB
|
| 3
|
30.018 dB
|
| 4
|
30.052 dB
|
Arithmetic Mean: 30.040250 dB
Standard Deviation: 0.018 dB
Standard Uncertainty: 9.132·10-3 dB
Degrees of Freedom: 3
|
MEASUREMENTS: Four observations are made of the incremental loss of the attenuator to be calibrated between settings of 0 dB and 30 dB. (see EAL-R2-S1:S7.11)
|
dLS: |
Type B normal distribution
Value: 0.003 dB
Expanded Uncertainty: 0.005 dB
Coverage Factor: 2
|
REFERENCE ATTENUATOR: The calibration certificate for the reference attenuator gives a value of attenuation for the 30,000 dB setting at 10 GHz of 30,003 dB with an associated expanded uncertainty of 0,005 dB (coverage factor k=2). The correction of +0,003 dB with the associated expanded uncertainty of 0,005 dB (coverage factor k=2) is considered to be valid for attenuation settings of the reference attenuator which differ not more ±0,1 dB from the calibrated setting of 30,000 dB.
|
dLD: |
Type B rectangular distribution
Value: 0.0 dB
Halfwidth of Limits: 0.002 dB
|
DRIFT OF THE REFERENCE: The drift of the attenuation of the reference attenuator is estimated from its calibration history to be zero with limits ±0,002 dB.
|
dLM: |
Type B U-shaped distribution
Value: 0.0 dB
Halfwidth of Limits: 0.0283 dB
|
MISMATCH LOSS: (see EAL-R2-S1:S7.6)
|
dLK: |
Type B rectangular distribution
Value: 0.0 dB
Halfwidth of Limits: 0.003 dB
|
LEAKAGE CORRECTION: Leakage signals through the attenuator to be calibrated have been estimated from the measurements at 0 dB setting to be at least 100 dB below the measurement signal. The correction for leakage signals is estimated from these findings to be within ±0,003 dB at the 30 dB setting.
|
dLib: |
Type B rectangular distribution
Value: 0.0 dB
Halfwidth of Limits: 0.0005 dB
|
RESOLUTION OF THE REFERENCE ATTENUATOR SETTING: The digital readout of the reference attenuator has a resolution of 0,001 dB from which the correction for resolution is estimated to be within ±0.0005 dB.
|
dLia: |
Type B rectangular distribution
Value: 0.0 dB
Halfwidth of Limits: 0.0005 dB
|
RESOLUTION OF THE REFERENCE ATTENUATOR SETTING: The digital readout of the reference attenuator has a resolution of 0,001 dB from which the correction for resolution is estimated to be within ±0.0005 dB.
|
dL0b: |
Type B normal distribution
Value: 0.0 dB
Expanded Uncertainty: 0.002 dB
Coverage Factor: 1.0
|
RESOLUTION OF THE NULL DETECTOR: The detector resolution was determined from a previous evaluation to have a standard deviation of 0,002 dB at each reading with assumed normal probability distribution.
|
dL0a: |
Type B normal distribution
Value: 0.0 dB
Expanded Uncertainty: 0.002 dB
Coverage Factor: 1
|
RESOLUTION OF THE NULL DETECTOR: The detector resolution was determined from a previous evaluation to have a standard deviation of 0,002 dB at each reading with assumed normal probability distribution.
Uncertainty Budgets:
LX: attenuation of the attenuator to becalibrated |
Quantity
|
Value
|
Standard Uncertainty
|
Distribution
|
Sensitivity Coefficient
|
Uncertainty Contribution
|
Index
|
|---|
| LS
|
30.040250 dB
|
9.132·10-3 dB
|
normal
|
1.0
|
9.1·10-3 dB
|
16.6 %
|
| dLS
|
3.000·10-3 dB
|
2.500·10-3 dB
|
normal
|
1.0
|
2.5·10-3 dB
|
1.2 %
|
| dLD
|
0.0 dB
|
1.155·10-3 dB
|
rectangular
|
1.0
|
1.2·10-3 dB
|
0.3 %
|
| dLM
|
0.0 dB
|
0.02001 dB
|
U-distr.
|
1.0
|
0.020 dB
|
79.7 %
|
| dLK
|
0.0 dB
|
1.732·10-3 dB
|
rectangular
|
1.0
|
1.7·10-3 dB
|
0.6 %
|
| dLib
|
0.0 dB
|
288.7·10-6 dB
|
rectangular
|
1.0
|
290·10-6 dB
|
0.0 %
|
| dLia
|
0.0 dB
|
288.7·10-6 dB
|
rectangular
|
-1.0
|
-290·10-6 dB
|
0.0 %
|
| dL0b
|
0.0 dB
|
2.000·10-3 dB
|
normal
|
1.0
|
2.0·10-3 dB
|
0.8 %
|
| dL0a
|
0.0 dB
|
2.000·10-3 dB
|
normal
|
-1.0
|
-2.0·10-3 dB
|
0.8 %
|
| LX
|
30.04325 dB
|
0.02242 dB
|
Results:
|
Quantity
|
Value
|
Expanded Uncertainty
|
Coverage factor
|
Coverage
|
|---|
| LX
|
30.043 dB
|
0.045 dB
|
2.00
|
95% (t-table 95.45%)
|
|